HandyLab 700 Portable pH meters
Features
- Contrast-rich and scratch-resistant clear glass LCD display
- Self-explanatory operation with one clear text line
- Display of the sensor status at one glance with Sensoface
- Calibration of up to 3 points with automatic buffer detection (from 10 buffers)
- Manual calibration by indicating random buffer values
- Real-time clock and display of battery charging status
- Extended operating time of more than 1,000 hours with standard batteries (4 x AA)
Measuring parameters | HandLab 700 | |
Memosens pH (ISFET) | pH Display area | -2.000 … +16.000 pH |
mV Display area | -2000 … +2000 mV | |
Temp. Display area | -50 … +250 °C | |
Memosens Redox | mV Display area | -2000 … +2000 mV |
Temp. Display area | -50 … +250 °C | |
pH/mV (Analog) | pH measure range | -2 … 16 |
mV measure range | -1300 … +1300 | |
Temperature | NTC 30 kΩ | |
Pt 1000 | ||
Autocal Buffer Sets | Knick CaliMat, Ciba (94), Hamilton, NIST Technical, HACH, Mettler-Toledo, NIST-Standard, SI-Analytics techn. buffer, DIN 19267, Reagecon |
|
Perm. calibration range | Zero point : | 6 … 8 pH |
ISFET: | -750 … +750 mV | |
Slope: | approx. 74 … 104 % | |
Calibration Timer | Sample interval 1 … 99 days, can be switched off | |
Display | LCD STN 7-segment display with three lines and symbols | |
Diagnostic Functions | Sensor data (Memosens only) Manufacturer, sensor type, serial number, operating time | |
Energy supply | Batteries 4x AA (Mignon) alkaline or lithium | |
Operating time | approx. 1,000 h (Alkaline) | |
Material | PA12GF30+TPE | |
Dimensions/Weight | 132 x 156 x 30 mm, 500 g | |
Type of protection | IP66/67 with pressure compensation | |
EMV | DIN EN 61326-1 (General Requirements) Interference Emission Class B (residential) Interference resistance Industry Branch DIN EN 61326-2-3 (Special Requirements for Pressure Transducers) |
|
RoHS Conformity | As per directive 2011/65/EU |